Difference between revisions of "Fluorometer"

From CleanEnergyWIKI
Jump to navigation Jump to search
Line 1: Line 1:
=== Background ===
=== Background ===
[[Image:Fluorometer_layout.png|thumb|400px|Internal layout of fluorometer components ]]
[[Image:Fluorometer_layout.png|thumb|400px|Internal layout of fluorometer components- Fluorolog-3, model FL3-12 (Horiba Jobin Yvon) ]]
A fluorometer consists of an excitation monochromator which directs light of specific wavelength on a sample. A fluorescent sample produces and emission spectrum that is measured through a monochromator connected to a photomultiplier detector similar to spectrophotometer.
A fluorometer consists of an excitation monochromator which directs light of specific wavelength on a sample. A fluorescent sample produces and emission spectrum that is measured through a monochromator connected to a photomultiplier detector similar to spectrophotometer.



Revision as of 08:32, 16 June 2011

Background

Internal layout of fluorometer components- Fluorolog-3, model FL3-12 (Horiba Jobin Yvon)

A fluorometer consists of an excitation monochromator which directs light of specific wavelength on a sample. A fluorescent sample produces and emission spectrum that is measured through a monochromator connected to a photomultiplier detector similar to spectrophotometer.

Significance

Excitation and fluorescence of sample in cuvette

Fluorescence quantum yield determination using relavitive method

One significant use of the fluorometer (or fluorimeter) is the determination of the fluorescence quantum yield. This is done using a relative method based on a reference compound of known quantum yield. The unknown sample and the reference sample are measured at the same excitation wavelengths and measurement conditions. The wavelength-integrated fluorescent intensity of both materials are then used in the calculation:

<math>\phi = \phi_{reference} \frac {F_{sample} / A_{sample}} {F_{reference}/ A_{reference}} \left( \frac {n_{sample}} {n_{reference}} \right) ^2\,\!</math>

where

φ is the quantum yield

F= integrated fluorescence intensity

A= absorbance at excitation wavelength

n= refractive index

Optically dilute solution

The intensity of the excitation beam should be almost constant along excitation beam. The fluorescence signal is proportional to intensity of excitation beam and the intensity has to be the same for sample and reference.

Typically: A ≤ 0.02 over 1 cm path length

From entrance face to center of cuvette: A = 0.01

T= 10-A = 10-0.01= 0.977

Therefore the intensity has changed only by 2%

However, depending on absorption spectrometer used, measurement of A in this range may not be accurate enough. How to proceed?

  • Measure A on higher concentration solution
  • Dilute solution by (accurately) known factor
  • Perform fluorescence measurement on diluted solutions

Choice of excitation wavelength

Excitation wavelength should be within the absorption band of the compounds

The excitation wavelength should be within the absorption band of the compounds. The same excitation wavelength is used for reference and sample compounds. The emission spectrum is collected on the long wavelength side of the excitation wavelength (to avoid strong scattered light from excitation beam).

In this test, we are using: λexc = 350 nm

Repeated measurements

Its always a good idea to have multiple data points! Prepare multiple dilutions and then measure fluorescence emission spectrum of each solution. Determine the slope of the line F/A for sample and reference. Deviations from linearity could indicate that emission was affected by reabsorption

Fluorescence intensity for various dilutions

Reabsorption

Reabsorption is the absorption of the emitted light by the same solution before light exits cuvette. This is more significant for compounds with small Stokes shifts. Reabsorption can appear as a redshift (or decrease in fluorescence intensity on the short wavelength portion of the spectrum). The effect can be minimized by reducing concentration of solution.
Corrected fluorescence spectra

Solvent subtraction.png

Detectors and gratings do not have the same efficiency at all wavelengths. Therefore the results need to be corrected by a factor that accounts for wavelength response of the instrument. The contribution of the solvent (Raman scattering) and noise (dark counts) should also be subtracted.

Sample Calculation

F/A values n

Sample (#1): 1.289 x 1010 cps/mA 1.3288 (methanol)

Reference: 1.316 x 1010 cps/mA 1.4266 (cyclohexane)

φreference = 0.87)

<math>\phi = \phi_{reference} \frac {F_{sample} / A_{sample}} {F_{reference}/ A_{reference}} \left( \frac {n_{sample}} {n_{reference}} \right) ^2\,\!</math>


<math>\phi = 0.87 * \frac {1.289} {1.316} * \left( \frac {1.3288} {1.4266} \right) ^2\,\!</math>

Operation

External Links

Wikipedia:Fluorescence spectroscopy