X-ray Photoelectron Spectroscopy and UV Photoelectron Spectroscopy are techniques for studying the surface characteristic of materials.
<table id="toc" style="width: 100%">
<tr>
<td style="text-align: center; width: 33%">[[Main_Page#Research Equipment, Devices and Techniques|Return to Research Tool Menu]]</td>


===What is the Problem?===
</tr>
</table>
 
 
X-ray Photoelectron Spectroscopy and UV Photoelectron Spectroscopy are techniques for studying surface characteristics of materials.
 
===Overview Physics of XPS and UPS===


OLEDs and OPVs consist of thin films of organic materials, sandwiched between contacting electrodes.  We need analytical tools which tell us:
OLEDs and OPVs consist of thin films of organic materials, sandwiched between contacting electrodes.  We need analytical tools which tell us:
*Elemental composition of metal, metal oxide and organic surfaces (top 1-10 nm)
*Elemental composition of metal, metal oxide and organic surfaces (top 1-10 nm)
*The molecular state of those elements in that same region
*The molecular state of those elements in that same region
*The frontier orbital energies which control rates of charge transfer, photopotentials, onset voltages, etc.
*The frontier orbital energies which control rates of charge transfer, photopotentials, onset voltages, etc. see [[Work Function of Metals]]
 


===What is our approach? Physics of XPS and UPS===
[[Image:Surface_electron_spectroscopies.jpg|thumb|right|300px |XPS uses high energy X-ray photons to excite “core” electrons in the near-surface region  UPS uses lower energy photons in the deep UV region to excite valence electrons.]]
[[Image:Surface_electron_spectroscopies.jpg|thumb|right|300px |XPS uses high energy X-ray photons to excite “core” electrons in the near-surface region  UPS uses lower energy photons in the deep UV region to excite valence electrons.]]
We use high-vacuum surface electron spectroscopies: X-ray photoelectron spectroscopy (XPS)and UV-photoelectron spectroscopy (UPS) to provide the elemental, molecular and energetic information we require about these materials.Surface analysis is carried out in high vacuum spectrometers, with sophisticated sample handling capabilities. The sample is prepared in a chamber to which a variety of devices can be attached.   The idea is to keep the surface as clean as possible, and to selectively add monolayers of organic materials to these surfaces, without the need to break vacuum between analyses.The sample is located at the center of the analytical chamber, and positioned so that we can excite it with either X-rays or UV photons.Once the photoelectrons are ejected from the sample, they are collected by a series of focusing lenses, and then separated according to their kinetic energy in a “hemispherical” analyzer.  We use either a single “channeltron” detector (UPS) or a multi-channel detector(XPS)
We use high-vacuum surface electron spectroscopies: X-ray photoelectron spectroscopy (XPS)and UV-photoelectron spectroscopy (UPS) to provide the elemental, molecular and energetic information we require about these materials. Surface analysis is carried out in high vacuum spectrometers, with sophisticated sample handling capabilities. The sample is prepared in a chamber to which a variety of devices can be attached. The idea is to keep the surface as clean as possible, and to selectively add monolayers of organic materials to these surfaces, without the need to break vacuum between analyses. The sample is located at the center of the analytical chamber, and positioned so that we can excite it with either X-rays or UV photons. Once the photoelectrons are ejected from the sample, they are collected by a series of focusing lenses, and then separated according to their kinetic energy in a “hemispherical” analyzer.  We use either a single “channeltron” detector (UPS) or a multi-channel detector(XPS)


===Technique ===
[[Image:PS-surfaceanalysis.jpg|thumb|right|300px |The small sampling depth of XPS and UPS arises because most of the photoelectrons generated do NOT make it out of the solid – they are scattered below the surface and not detected.  Only those within 1-10 nm of the surface get out and can be analyzed.]]


[[Image:PS-surfaceanalysis.jpg|thumb|right|300px |The small sampling depth of XPS and UPS arises because most of the photoelectrons generated do NOT make it out of the solid – they are scattered below the surface and not detected.  Only those within 1-10 nm of the surface get out and can be analyzed.]]
{{#ev:youtube|kK5OBNJwnws}}


<swf width= "640" height="480">http://depts.washington.edu/cmditr/media/pes.swf</swf>
Other XPS examples
{{#ev:youtube|pTLJ2zW_z6c}}
XPS Analysis
{{#ev:youtube|v1dyfwHdvbA}}


<br clear="all">
<br clear="all">


===Data interpretation===
===Significance===
<gallery widths=300px heights=200px perrow=3> Image:Xps_spectrum.png‎|Here is a typical “survey spectrum” for a clean gold surface, modified with a single monolayer of a phenyl-terminated alkanethiol.  We plot number of emitted photoelectrons on the y-axis, and their binding energy (BE) or kinetic energy (KE) on the x-axis.  The photoelectric effect applies (as first described by Einstein) – there is conservation of energy, and you can see that the BE of each photoelectron is related to the excitation source energy, the measured kinetic energy of that photoelectron, and a “work function” of the spectrometer (constant).  Each element provides us with at least one photoemission peak, with a distinct BE or KE.  
<gallery widths=300px heights=200px perrow=3> Image:Xps_spectrum.png‎|Here is a typical “survey spectrum” for a clean gold surface, modified with a single monolayer of a phenyl-terminated alkanethiol.  We plot number of emitted photoelectrons on the y-axis, and their binding energy (BE) or kinetic energy (KE) on the x-axis.  The photoelectric effect applies (as first described by Einstein) – there is conservation of energy, and you can see that the BE of each photoelectron is related to the excitation source energy, the measured kinetic energy of that photoelectron, and a “work function” of the spectrometer (constant).  Each element provides us with at least one photoemission peak, with a distinct BE or KE.  


Line 34: Line 47:
Image:Ups_fluoridated_alkanes.jpg|Now we see a series of UPS data for Au surface modified with alkanethiols of different lengths (from C3 to C18) (left panel) and a similar series of alkanethiols which are fluorinated at 1,2,4 and 10 positions along the chain (right panel).  At the bottom of the screen we have schematically shown the orientation of the molecular dipole moments for these modifiers – the normal alkanes point the positive end of the dipole away from the surface, and lower the work function; the fluorinated alkanes point the negative end of the dipole away from the surface and increase the work function.
Image:Ups_fluoridated_alkanes.jpg|Now we see a series of UPS data for Au surface modified with alkanethiols of different lengths (from C3 to C18) (left panel) and a similar series of alkanethiols which are fluorinated at 1,2,4 and 10 positions along the chain (right panel).  At the bottom of the screen we have schematically shown the orientation of the molecular dipole moments for these modifiers – the normal alkanes point the positive end of the dipole away from the surface, and lower the work function; the fluorinated alkanes point the negative end of the dipole away from the surface and increase the work function.


Image:Workfunction_dipole.png|The presence of these surface modifiers either increases or decrreases the energy of the low kinetic energy edge of the UV-photoemission spectrum, and these shifts correspond to changes in the effective surface work function of the conductor.  Here we show that we can plot these changes in effective work function as a function of the molecular dipole moment of the modifier.  The black squares correspond to self-assembled monolayers (SAMs) on gold (Au) surfaces, the red circles correspond to similar series of molecules on silver (Ag) surfaces, and the blue triangles correspond to a series of phosphonic acids attached to indium-tin oxide (ITO) surfaces.  Note that the slopes of these plots are nearly the same, suggesting that the molecular dipole moment for the modifier is the most important attribute of these molecules.  Note also that we can change the effective work function of these surfaces by up to ca. 1.8 eV!!
Image:Workfunction_dipole.png|The presence of these surface modifiers either increases or decreases the energy of the low kinetic energy edge of the UV-photoemission spectrum, and these shifts correspond to changes in the effective surface work function of the conductor.  Here we show that we can plot these changes in effective work function as a function of the molecular dipole moment of the modifier.  The black squares correspond to self-assembled monolayers (SAMs) on gold (Au) surfaces, the red circles correspond to similar series of molecules on silver (Ag) surfaces, and the blue triangles correspond to a series of phosphonic acids attached to indium-tin oxide (ITO) surfaces.  Note that the slopes of these plots are nearly the same, suggesting that the molecular dipole moment for the modifier is the most important attribute of these molecules.  Note also that we can change the effective work function of these surfaces by up to ca. 1.8 eV!!
 
 


</gallery>
</gallery>
[[category:Research equipment]]

Latest revision as of 10:32, 18 January 2012

Return to Research Tool Menu


X-ray Photoelectron Spectroscopy and UV Photoelectron Spectroscopy are techniques for studying surface characteristics of materials.

Overview Physics of XPS and UPS

OLEDs and OPVs consist of thin films of organic materials, sandwiched between contacting electrodes. We need analytical tools which tell us:

  • Elemental composition of metal, metal oxide and organic surfaces (top 1-10 nm)
  • The molecular state of those elements in that same region
  • The frontier orbital energies which control rates of charge transfer, photopotentials, onset voltages, etc. see Work Function of Metals


XPS uses high energy X-ray photons to excite “core” electrons in the near-surface region UPS uses lower energy photons in the deep UV region to excite valence electrons.

We use high-vacuum surface electron spectroscopies: X-ray photoelectron spectroscopy (XPS)and UV-photoelectron spectroscopy (UPS) to provide the elemental, molecular and energetic information we require about these materials. Surface analysis is carried out in high vacuum spectrometers, with sophisticated sample handling capabilities. The sample is prepared in a chamber to which a variety of devices can be attached. The idea is to keep the surface as clean as possible, and to selectively add monolayers of organic materials to these surfaces, without the need to break vacuum between analyses. The sample is located at the center of the analytical chamber, and positioned so that we can excite it with either X-rays or UV photons. Once the photoelectrons are ejected from the sample, they are collected by a series of focusing lenses, and then separated according to their kinetic energy in a “hemispherical” analyzer. We use either a single “channeltron” detector (UPS) or a multi-channel detector(XPS)

Technique

The small sampling depth of XPS and UPS arises because most of the photoelectrons generated do NOT make it out of the solid – they are scattered below the surface and not detected. Only those within 1-10 nm of the surface get out and can be analyzed.

Other XPS examples

XPS Analysis


Significance