Difference between revisions of "Fluorometer"

From CleanEnergyWIKI
Jump to navigation Jump to search
Line 25: Line 25:
Intensity has to be the same for sample and reference
Intensity has to be the same for sample and reference
Typically: A ≤ 0.02 over 1 cm pathlength
Typically: A ≤ 0.02 over 1 cm pathlength
From entrance face to center of cuvette: �A = 0.01
From entrance face to center of cuvette: A = 0.01


Intensity has changed only by 2%
Intensity has changed only by 2%
Line 37: Line 37:
Excitation wavelength should be within the absorption band of the compounds
Excitation wavelength should be within the absorption band of the compounds
Same excitation wavelength to be used for reference and sample compounds
Same excitation wavelength to be used for reference and sample compounds
Emission spectrum collected �on the long wavelength side �of the excitation wavelength (to avoid strong scattered �light from excitation beam)
Emission spectrum collected on the long wavelength side of the excitation wavelength (to avoid strong scattered light from excitation beam)
In this test, we are using: �lexc = 350 nm
In this test, we are using: &lambda;<sub>exc</sub> = 350 nm
'''Repeated measurements'''
'''Repeated measurements'''
Prepare multiple dilutions
Prepare multiple dilutions
Line 48: Line 48:
'''Reabsorption'''
'''Reabsorption'''
Absorption of the emitted light by the same solution before light exits cuvette
Absorption of the emitted light by the same solution before light exits cuvette
More significant for compounds �with small Stokes shifts
More significant for compounds with small Stokes shifts
Reabsorption can appear �as a redshift (or decrease �in fluorescence intensity �on the short wavelength �portion of the spectrum)  
Reabsorption can appear as a redshift (or decrease in fluorescence intensity on the short wavelength portion of the spectrum)  
Effect can be minimized �by reducing concentration �of solution
Effect can be minimized by reducing concentration of solution
'''Corrected fluorescence spectra'''
'''Corrected fluorescence spectra'''
Detectors and gratings do not have the same efficiency at all wavelengths
Detectors and gratings do not have the same efficiency at all wavelengths
Results need to be corrected �by a factor that accounts �for wavelength response �of the instrument
Results need to be corrected by a factor that accounts for wavelength response of the instrument
The contribution of the �solvent (Raman scattering) �and noise (dark counts) �should also be subtracted
The contribution of the solvent (Raman scattering) and noise (dark counts) should also be subtracted
'''Sample Calculation'''
'''Sample Calculation'''
F/A values                      n
F/A values                      n
Sample (#1): 1.289 x 1010 cps/mA          1.3288 (methanol)
Sample (#1): 1.289 x 1010 cps/mA          1.3288 (methanol)
Reference: 1.316 x 1010 cps/mA   1.4266 (cyclohexane)
Reference: 1.316 x 1010 cps/mA   1.4266 (cyclohexane)
(freference = 0.87)
&phi;<sub>reference</sub> = 0.87)
 
:<math>\phi  =  \phi_{reference} \frac {F_{sample} / A_{sample}} {F_{reference}/ A_{reference}} \left( \frac {n_{sample}} {n_{reference}} \right) ^2\,\!</math>


:<math>\phi  =  0.87 * \frac {1.289} {1.316} * \left( \frac {1.3288} {1.4266} \right) ^2\,\!</math>
:<math>\phi  =  0.87 * \frac {1.289} {1.316} * \left( \frac {1.3288} {1.4266} \right) ^2\,\!</math>
:<math>\phi  =  \phi_{reference} \frac {F_{sample} / A_{sample}} {F_{reference}/ A_{reference}} \left( \frac {n_{sample}} {n_{reference}} \right) ^2\,\!</math>


=== Operation ===
=== Operation ===

Revision as of 10:19, 28 February 2011

Background

Significance

Fluorescence quantum yield determination using relavitive method

One significant use of the fluorometer (or fluorimeter) is the determination of the fluorescence quantum yield. This is done using a relative method based on a reference compound of known quantum yield. The unknown sample and the reference sample are measured at the same excitation wavelengths and measurement conditions. The wavelength-integrated flourescent intensity of both materials are then used in the calculation:

<math>\phi = \phi_{reference} \frac {F_{sample} / A_{sample}} {F_{reference}/ A_{reference}} \left( \frac {n_{sample}} {n_{reference}} \right) ^2\,\!</math>

where

φ is the quantum yield

F= integrated fluorescence intensity

A= absorbance at excitation wavelength

n= refractive index

Optically dilute solution Intensity of excitation beam should be almost constant along excitation beam Fluorescence signal is proportional to intensity of excitation beam Intensity has to be the same for sample and reference Typically: A ≤ 0.02 over 1 cm pathlength From entrance face to center of cuvette: A = 0.01

Intensity has changed only by 2% Depending on absorption spectrometer used, measurement of A in this range may not be accurate enough How to proceed? Measure A on higher concentration solution Dilute solution by (accurately) known factor Perform fluorescence measurement on diluted solutions

Choice of excitation wavelength Excitation wavelength should be within the absorption band of the compounds Same excitation wavelength to be used for reference and sample compounds Emission spectrum collected on the long wavelength side of the excitation wavelength (to avoid strong scattered light from excitation beam) In this test, we are using: λexc = 350 nm Repeated measurements Prepare multiple dilutions Measure fluorescence emission spectrum of each solution Determine the slope of the line F/A for sample and reference Always a good idea to have multiple data points!

Deviations from linearity could indicate that emission was affected by reabsorption Reabsorption Absorption of the emitted light by the same solution before light exits cuvette More significant for compounds with small Stokes shifts Reabsorption can appear as a redshift (or decrease in fluorescence intensity on the short wavelength portion of the spectrum) Effect can be minimized by reducing concentration of solution Corrected fluorescence spectra Detectors and gratings do not have the same efficiency at all wavelengths Results need to be corrected by a factor that accounts for wavelength response of the instrument The contribution of the solvent (Raman scattering) and noise (dark counts) should also be subtracted Sample Calculation F/A values n Sample (#1): 1.289 x 1010 cps/mA 1.3288 (methanol) Reference: 1.316 x 1010 cps/mA 1.4266 (cyclohexane) φreference = 0.87)

<math>\phi = \phi_{reference} \frac {F_{sample} / A_{sample}} {F_{reference}/ A_{reference}} \left( \frac {n_{sample}} {n_{reference}} \right) ^2\,\!</math>
<math>\phi = 0.87 * \frac {1.289} {1.316} * \left( \frac {1.3288} {1.4266} \right) ^2\,\!</math>

Operation

External Links

Wikipedia:Fluorescence spectroscopy